MCTBIO worst-case kit: unscrew implants by reverse torque (3 fixture removers, wrench, 7 trephines) and remove fractured screws via guide sleeves. Universal for common implant systems.

COMPLICATION MANAGEMENT · MCT BIO
MCTBIO Implant and Screw Removal Kit
The worst-case kit for both emergencies in one box: unscrew failed implants by reverse torque – combined with a trephine if needed – and remove fractured abutment and prosthetic screws via guide sleeves. Universal for common implant systems.
IMPLANT
3 fixture removers
mini · regular · wide
TREPHINES
7 pieces
for difficult cases
SCREW
guide sleeves
drill and tap
REVERSE TORQUE
up to 200 Ncm
per manufacturer
SYSTEMS
universal
internally threaded implants
PART 1 · REMOVE IMPLANT
Unscrew instead of drilling out

The fixture remover is turned counterclockwise into the implant's internal thread until it wedges. With the flipped wrench, the implant is then released from the bone by reverse torque – the bony walls are largely preserved.

The three sizes fit any implant with an internal thread. Seven trephines are included for particularly well-integrated implants.

Fixture remover Implant Ø Note
Mini 3–4 mm narrow implants
Regular 4–5 mm standard
Wide 5–8 mm wide implants
A smaller remover also engages larger implants but can be hard to release afterwards.
MCT BIO implant and screw removal kit – box with fixture removers, trephines and wrench
The complete kit
WORKFLOW
Unscrewing the implant

1. Remove prosthetics and abutment screw, clean the internal thread.
2. Choose the fixture remover matching the implant diameter.
3. Turn the remover counterclockwise into the internal thread with the hand adapter until it wedges; confirm the seat with the wrench.
4. Remove the wrench, flip it and unscrew the implant with controlled reverse torque.
5. After removal, release the remover from the implant clockwise with the wrench.

COMBINATION TECHNIQUE
If the implant does not give at 200 Ncm
More reverse torque is not recommended. Instead, use the matching trephine for only 2–3 mm or about one third to one half of the implant length – never to the apex – and then unscrew the implant with fixture remover and wrench. This preserves considerably more bone than full trephination.
PART 2 · REMOVE SCREW
Safely retrieving fractured screws

The guide sleeve is placed on the implant connection and centres the remover drill exactly on the fragment, so the drill cannot touch the internal thread.

The drill cuts a small dimple into the fragment in reverse, and the reverse tap grips it and unscrews the fragment.

Step Speed Torque Direction
Remover drill 1,000–1,250 rpm 50–70 Ncm reverse
Tap 70–80 rpm 50–70 Ncm reverse
Guide values according to the technique guide; use a reduction contra-angle.
Guide sleeves for different implant connections
Guide sleeves
WORKFLOW
Removing the screw

1. Place the matching guide sleeve and stabilise it with the handle.
2. With the remover drill in reverse and an up-and-down motion, drill a 1–2 mm dimple into the fragment – with external irrigation, suction shavings.
3. Insert the tap, low speed in reverse.
4. Guide the tap into the dimple, unscrew the fragment, check the thread with a new screw.

COMPATIBILITY
For which connections?
Guide sleeves are available for all common connection types, e.g.:
Connection Examples
External hex Brånemark type, 3i External, Hiossen US, MegaGen ExFeel External
Internal hex Dentium, Hiossen SS/GS, MegaGen, Neobiotech
Conical NobelActive, Straumann Bone Level, Astra Tech
Internal octagon / tissue level Straumann Tissue Level, SwissPlus
TSV type Zimmer TSV, BioHorizons, MIS, Implant Direct Legacy
Tri-lobe Nobel Replace, Implant Direct RePlant
Please enquire which guide sleeves are included in the kit; further sleeves available individually. Brand names belong to their respective owners and are used for identification only.
PRACTICAL TIPS
What matters
 
  • Determine implant system and diameter before the procedure (implant card, radiograph).
  • Increase reverse torque slowly – some patients find high torque painful.
  • Never take the trephine to the implant apex; irrigate copiously.
  • After screw removal, clarify the cause of the fracture and always use a new screw.
Note: For use by trained dental professionals only. Use, cleaning and sterilisation according to the manufacturer's instructions. Successful removal cannot be guaranteed in every case.